Multicenter comparative study of three “non-destructive” methods of detecting the cleanliness of the da Vinci surgical robotic instrument
Original Article

Multicenter comparative study of three “non-destructive” methods of detecting the cleanliness of the da Vinci surgical robotic instrument

Aiqin Chen1#, Xiumei Zou2#, Yanhua Tan3, Yanhan Chen1, Xinru Ye1, Shuqin Hao4

1Central Sterile Supply Department, Sun Yat-sen University Cancer Center, Guangzhou, China; 2Intensive Care Unit, Sun Yat-sen University Cancer Center, Guangzhou, China; 3Central Sterile Supply Department, First Affiliated Hospital of Guangzhou Medical University, Guangzhou, China; 4Central Sterile Supply Department, Shenzhen Second People’s Hospital, Shenzhen, China

Contributions: (I) Conception and design: S Hao; (II) Administrative support: S Hao; (III) Provision of study materials or patients: A Chen, Y Tan; (IV) Collection and assembly of data: A Chen, Y Tan; (V) Data analysis and interpretation: A Chen, Y Tan; (VI) Manuscript writing: All authors; (VII) Final approval of manuscript: All authors.

#These authors contributed equally to this work.

Correspondence to: Shuqin Hao. Shenzhen Second People’s Hospital, Shenzhen 518020, China. Email: 1059893164@qq.com.

Background: The robotic instrument of the da Vinci surgical system determines the accuracy of robotic-assisted surgery, However, the most effective cleaning method of robotic equipment is a challenge for healthcare professionals. This study compared three da Vinci robot-assisted surgery manipulators to detect the effect of “non-destructive” testing of the cleaning effect by two different methods.

Methods: The post-surgical cleaning of the da Vinci robotic instruments in the Sun Yat-sen University Cancer Center, The First Affiliated Hospital of Guangzhou Medical University and the Shenzhen Second People’s Hospital was performed using two different processes from January 2019 to January 2020: manual joint automatic ultrasonic cleaning and automatic mechanical cleaning. The efficacy of visual estimation, the residual protein assay (quantitative) and adenosine triphosphate (ATP) biological biofluorescence detection of the cleanliness of the mechanical instrument’s work (distal working end) surface and the shaft’s inner chamber was compared. If the cleaning effect of any position on the surface or inner cavity of the manipulator did not qualify, the entire robotic instrument was judged as disqualified.

Results: A total of 198 cases of da Vinci robotic instrument postoperative cleanliness data were collected. The qualifying rates of automatic ultrasonic cleaning were 96.97% by visual estimation, 93.94% by residual protein assay and 60.61% by ATP biological fluorescence detection. The respective rates for automatic mechanical cleaning were 100% by visual estimation, 90.91% by residual protein assay and 66.67% by ATP biological fluorescence detection.

Conclusions: The cleaning of the da Vinci robotic instrument detected by “non-destructive” residual protein assay or ATP biological fluorescence detection is more accurate than visual estimation.

Keywords: Adenosine triphosphate (ATP) assay; cleanliness testing; Da Vinci robotic surgical instruments; residual protein assay; visual estimation


Submitted Oct 12, 2021. Accepted for publication Dec 10, 2021.

doi: 10.21037/gs-21-814


Introduction

In 2000, the da Vinci robotic surgical system was approved by the Food and Drug Administration in the US (1), and since Horgan first reported using it in an adrenalectomy in 2001 (2), robotic-assisted surgery has been increasingly used. In the past decade, robot-assisted surgery with the da Vinci surgical system has increased rapidly in mainland China (3), and the average number of procedures per robot is far ahead of that in other developed countries (4).

The robotic instrument is the high-value consumable of the da Vinci system that both determines the accuracy of robotic-assisted surgery and limits the number of times the system can be used. It is usually designed as a long tubular structure with a semi-blind end and many gaps and axial joints at the working end (5). Because of the complex structure of robotic instruments, their cleaning is a challenge for healthcare staff (6). The robotic instrument in particular is more likely to have a large amount of blood residue after surgery, and thus cleaning is even more difficult than for the other mechanical parts of the da Vinci robot. Therefore, the aim of the present study was how to judge if the equipment’s postoperative cleanliness has achieved the standard of clinical use.

The da Vinci working group conducted a comparative study of the cleaning effect of the working end of da Vinci Maryland bipolar forceps (7), tested by either “destructive” or “non-destructive” methods, based on a quantitative protein assay. The comparisons were conducted under laboratory conditions using professional eluent shakers and other specific tools. However, the method adopted in that study is not easy to use in the actual work of hospital sterile supply centers, and some steps even require the destruction of instruments for sampling, which is difficult to apply in daily clinical work. Moreover, there is no authoritative detection guideline for evaluating the cleaning effect of robotic assistants in surgery, especially for robotic instruments, and the actual cleaning operation is only performed in accordance with the guidelines of the device manufacturer.

In this multicenter study, to provide a scientific reference for the management of da Vinci surgical instruments, we evaluated the reliability of the cleaning process of the robotic instrument, as well as the advantages and disadvantages of three common methods of detecting the cleanliness of surgical instruments (8,9), namely, visual estimation, quantitative protein assay, and adenosine triphosphate (ATP) biological fluorescence detection. We present the following article in accordance with the STROBE reporting checklist. We present the following article in accordance with the STROBE reporting checklist (available at https://dx.doi.org/10.21037/gs-21-814).


Methods

Study subjects

The da Vinci surgical robotic instruments requiring postoperative cleaning at the Sun Yat-sen University Cancer Center, the First Affiliated Hospital of Guangzhou Medical University and Shenzhen Second People’s Hospital were sampled from January 2019 to January 2020.

Cleaning process

Manual combined ultrasonic cleaning

The specific steps for the da Vinci robot instrument were: (I) pretreatment (initial brushing with a nylon brush under flowing water); (II) rinsing (pressure rinsing of the main rinse port and other rinse ports for at least 20 s); (III) ultrasonic cleaning (injection of ≥15 mL of multienzyme solution, complete submersion in the cleaning solution, ultrasonic cleaning for 15 min); (IV) rinse again (pressure rinse the main rinse port and other rinse ports for at least 20 s); (V) brush again (with a clean nylon brush, under running water); (VI) rinse again (rinse the outer surface thoroughly for 60 s); (VII) dehydration (keep the clamping end of the robot instrument raised until there is no residual water, then each rinse port is blown dry with compressed air).

Fully automated mechanical cleaning

A dedicated cleaning program without the moist heat disinfection step was set up in accordance with the washing and disinfection instructions specified by manufacturer. Specifically, after the robot instrument is retrieved, it is disassembled to the smallest unit and visible blood flushed away (dried blood stains first soaked in detergent and brushed); the equipment is loaded into the dedicated cleaning layer rack (water injection port 1 and water injection port 2 connected properly with the da Vinci dedicated layer rack); the special test program is commenced to complete the flush, wash, rinse, final rinse and dry operations.

Disinfection steps for the two cleaning processes were not included, to avoid affecting the ATP biological fluorescence detection results.

Testing order and grouping method

According to the order in which the robotic instruments were removed after cleaning, they were divided into a total of six groups: manual combined ultrasonic cleaning with visual estimation, automatic mechanical cleaning with visual estimation; manual combined ultrasonic cleaning with residual protein assay, automatic mechanical cleaning group with residual protein assay; manual combined ultrasonic cleaning with ATP biological fluorescence detection, and automatic mechanical cleaning with ATP detection.

Detection and sampling methods

A “non-destructive” sampling method was adopted; that is, the most difficult cleaning position of the instrument was sampled without damaging the robotic instrument.

Each robotic instrument was cleaned immediately after recovery, and the cleanliness of the surface of the working end of the robotic instrument and the inner cavity of the bar shaft were tested by the three methods within 1 h after the working end had been cleaned.

If any surface of the working end of each robot instrument or any position in the inner cavity for the bar shaft did not qualify, the cleaning of the entire instrument was deemed to be disqualified.

Visual estimation

Surface of working end

The tooth space, the cable and pulley gap at the working end of the instrument were comprehensively examined by a magnifying glass (×3–5). The active gap and the cavity of the instrument were tested by wiping with a white and non-shedding cotton shavings of cotton swab. Any dirt residue was deemed as disqualification (Figure 1).

Figure 1 Detecting the cleaning effect of the da Vinci surgical robotic instrument (visual estimation).
Inner cavity of the bar shaft

The cleanliness of the inner cavity was initially tested by a “swirling water method”: the two ends of the robot instrument are held in the hands, shaking it up and down with the water injection port downward, then flushing the residual liquid in the inner cavity of the bar shaft onto a white absorbent towel laid out in advance and observing the color of the residual liquid. Any dirt residue was deemed as disqualification. To avoid any residue of a special nature from the inner cavity structure, the cleanliness test of the cleaning rod was excluded (Figure 1).

Residual protein assay

Surface of working end

After the visual estimation test had been passed, the tooth space of the working end and the gap of the shaft joint were wiped with a protein residue testing swab (after wetting with purified water), after turning the three control knobs to fully expose the shaft joint and gap so that the equipment could be fully sampled. Each position was repeatedly wiped three times by testing swab, then fully mixed with the neutralizer and placed in the culture reader to read the results (Figure 2).

Figure 2 Detecting the cleaning effect of the da Vinci surgical robotic instrument (residual protein assay).
Inner cavity of the bar shaft

After the visual estimation test had been passed, 10 mL of sterile injection water was injected to elute in water injection port 1, followed by the water injection port facing downward, so that the injected eluate discharged into a sterile bowl. Air was injected repeatedly three times with a syringe to fully discharge the residual eluate. The eluate was extracted from the bowl and the process repeated twice. The final eluate was sampled by protein residue testing swab and neutralized with the neutralizing agent. The results were read by a special reader after mixing thoroughly with the neutralizing agent attached to the testing swabs (Figure 2).

ATP biological biofluorescence detection

Surface detection of working end

After the visual estimation test was passed, an ATP surface sampling swab moistened with sterile injection water was wiped over the surface of the working end, the tooth space, and cable gap, after the three control knobs had been turned to fully expose the axial joint and gap parts. Each part was wiped three times, then cultured and the results were read in accordance with the operation instructions (Figure 3).

Figure 3 Detecting the cleaning effect of the da Vinci surgical robotic instrument (ATP biological biofluorescence detection). ATP, adenosine triphosphate.
Internal cavity of the bar shaft

The same procedure as for the residual protein assay was performed and the final eluate was sampled with an ATP biofluorescence sampling swab and neutralized with the neutralizer. The results were read by a special reader after mixing thoroughly with the neutralizing agent attached to the testing swabs (Figure 3).

Statistical analysis

Statistical analysis was conducted by SPSS 20.0 software. The study data were quantitative, and comparison between the cleaning and detection methods was performed by chi-square test, and when performed by paired chi-square test, P<0.05 was considered as a statistically significant difference.


Results

Comparison of fully automatic mechanical cleaning and manual combined ultrasonic cleaning

There was no statistical difference among visual estimation, residual protein assay and ATP biological fluorescence detection applied in the detection of manual combined ultrasound cleaning and fully automatic mechanical cleaning, all P>0.05. However, there were statistical differences in the comparison of the three detection methods applied to the two cleaning methods respectively, all P values <0.01 (Table 1).

Table 1

Comparison of overall qualifying rate of cleaning by three detection methods for two cleaning methods

Cleaning method Visual estimation Residual protein assay ATP biological fluorescence detection χ2 value P value
Manual combined ultrasound cleaning 96.97% (32/33) 93.94% (31/33) 60.61% (20/33) 45.06 <0.01
Fully automatic mechanical cleaning 100.00% (33/33) 90.91% (30/33) 66.67% (22/33) 39.53 <0.01
χ2 value 1.02 0.22 0.26
P value 0.31 0.64 0.61

ATP, adenosine triphosphate.

Comparison of visual estimation and residual protein assay

The results of the paired chi-square test suggested there was no statistical difference between visual estimation and the residual protein assay for the cleanliness of the da Vinci robot instrument (χ2 value =2.61, P>0.05) (Table 2).

Table 2

Comparison of qualifying rate of cleaning between visual estimation and residual protein assay

Visual estimation Residual protein assay
Unqualified Qualified In total
Unqualified 1 0 1
Qualified 4 61 65
In total 5 61 66

χ2 value =2.61, P=0.11.

Comparison of visual estimation and ATP biological fluorescence detection

The results of the paired chi-square test suggested there was no statistical difference between visual estimation and the ATP biological fluorescence detection for the cleanliness of the da Vinci robot instrument (χ2 value =0.08, P>0.05) (Table 3).

Table 3

Comparison of qualifying rate of cleaning between visual estimation and ATP biological fluorescence detection

Visual estimation ATP biological fluorescence detection
Unqualified Qualified In total
Unqualified 1 0 1
Qualified 23 42 65
In total 24 42 66

χ2 value =0.08, P=0.78. ATP, adenosine triphosphate.

Comparison of ATP biological fluorescence detection and residual protein assays

The results of the paired chi-square test suggested that there was a statistical difference between the results of the ATP biological fluorescence detection and the residual protein assay for the cleanliness of the da Vinci robot instrument (χ2 value=6.73, P<0.05) (Table 4).

Table 4

Comparison of qualifying rate of cleaning between ATP biological fluorescence detection and residual protein assay

Residual protein assay Visual estimation
Unqualified Qualified In total
Unqualified 5 0 5
Qualified 19 42 61
In total 24 42 66

χ2 value =6.73, P=0.009. ATP, adenosine triphosphate.


Discussion

Robotic-assisted surgery with the da Vinci system is being performed worldwide due to its utility and safety (10), but limited information is available on the effectiveness of cleaning robotic surgical instruments (6). Our comparison of visual estimation, residual protein assay (11) and ATP biological fluorescence detection (12,13), based on multicenter data to determine the cleaning effectiveness of the da Vinci robotic instrument, found that the “non-destructive” residual protein assay and the ATP biological fluorescence detection were more accurate than visual estimation.

The cleanliness of the working surface and internal cavity of the da Vinci robot instrument was detected by “non-destructive” methods in this study, from which we obtained results that were consistent with those from a “destructive” method (14). Others have also conducted research on detecting the cleaning effect of the da Vinci robotic instrument by a “non-destructive” method (15) that was similar to the residual protein sampling method used in our study. The positions in the robotic instrument that are difficult to clean can be sampled by the “non-destructive” sampling methods without damaging the instrument, which makes them suitable for preparing the robotic instruments for reuse.

The results of this study also indicated that whether manual cleaning or mechanical cleaning is applied, the qualifying rates for the cleanliness of the robotic instrument detected by visual estimation can achieve >95%, and the rate for the residual protein assay can also reach >90%, which indicates that either of these two methods can be used to reliably detect good cleaning results with standardized cleaning processes.

Visual estimation is an economical and easy-to-use method of testing for cleanliness. However, it can only observe contaminants >50 µg on the surface or in superficial positions of the instrument, and has difficulty in observe contaminants that are scattered, fine or in the lumen (8). For instruments with complex structures such as the robotic instrument of the da Vinci machine with its semi-blind end lumen (6), visual estimation alone is not comprehensive enough to detect the cleaning effect, which is also consistent with the results of this study. Therefore, whenever a new cleaning process is established, the cleaning process is changed, or there is a significant decrease in cleaning quality, we recommend using visual estimation combined with qualitative or quantitative testing methods such as the residual protein assay or other blood residue detection to “non-destructively” test the inner cavity and working end of the shaft to fully verify the effectiveness of the cleaning process, and this can also be used as one of the methods to verify the quality of periodic cleaning of the da Vinci robotic instrument.

The qualifying rates with ATP biological fluorescence detection in the study were too low, which we considered may have been due to the disinfection step being excluded in the cleaning procedure in this study (A0 value =600). The relative light unit released by the conversion of ATP in living cells is detected by ATP mainly with many interference factors. Studies have also shown that ATP is more suitable for the detection of the cleanliness of flexible endoscopy equipment (16). The blood residues on surgical instruments are mainly proteins, and the residual protein assay may be more suitable for the detection of cleaning effects of robotic instruments than visual estimation and ATP detection methods. We acknowledge the following shortcomings in the study. First, the sample size was small, mainly because of the high cost of the equipment and so the number of patients who undergo this type of surgery is not comparable with the number of patients who undergo conventional surgery. Second, only the comparison of the three different methods of detecting cleanliness was discussed in this study without evaluation of their effect on the postoperative prognosis of patients, which could be followed up with related studies. Finally, this was only a retrospective observational study with many confounding factors, so further studies are required to confirm the results.


Conclusions

It is recommended that the cleaning effect of the da Vinci robotic instrument should be regularly measured by “non-destructive” residual protein assay, ATP biological fluorescence detection and other quantitative detection methods in the inner shaft cavity and working end of the robotic instrument.


Acknowledgments

Funding: None.


Footnote

Reporting Checklist: The authors have completed the STROBE reporting checklist. Available at https://dx.doi.org/10.21037/gs-21-814

Data Sharing Statement: Available at https://dx.doi.org/10.21037/gs-21-814

Conflicts of Interest: All authors have completed the ICMJE uniform disclosure form (available at https://dx.doi.org/10.21037/gs-21-814). The authors have no conflicts of interest to declare.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Open Access Statement: This is an Open Access article distributed in accordance with the Creative Commons Attribution-NonCommercial-NoDerivs 4.0 International License (CC BY-NC-ND 4.0), which permits the non-commercial replication and distribution of the article with the strict proviso that no changes or edits are made and the original work is properly cited (including links to both the formal publication through the relevant DOI and the license). See: https://creativecommons.org/licenses/by-nc-nd/4.0/.


References

  1. Mak TW, Lee JF, Futaba K, et al. Robotic surgery for rectal cancer: A systematic review of current practice. World J Gastrointest Oncol 2014;6:184-93. [Crossref] [PubMed]
  2. Horgan S, Vanuno D. Robots in laparoscopic surgery. J Laparoendosc Adv Surg Tech A 2001;11:415-9. [Crossref] [PubMed]
  3. Zhou Q, Huang J, Pan F, et al. Operative outcomes and long-term survival of robotic-assisted segmentectomy for stage IA lung cancer compared with video-assisted thoracoscopic segmentectomy. Transl Lung Cancer Res 2020;9:306-15. [Crossref] [PubMed]
  4. Luo D, Liu Y, Zhu H, et al. The MicroHand S robotic-assisted versus Da Vinci robotic-assisted radical resection for patients with sigmoid colon cancer: a single-center retrospective study. Surg Endosc 2020;34:3368-74. [Crossref] [PubMed]
  5. Moran ME. Evolution of robotic arms. J Robot Surg 2007;1:103-11. [Crossref] [PubMed]
  6. Saito Y, Yasuhara H, Murakoshi S, et al. Challenging Residual Contamination of Instruments for Robotic Surgery in Japan. Infect Control Hosp Epidemiol 2017;38:143-6. [Crossref] [PubMed]
  7. Wehrl M, Michels W. A method for testing the cleaning of MIS robotic instruments. Central Service 2013;3:202-7.
  8. Lipscomb IP, Sihota AK, Keevil CW. Comparison between visual analysis and microscope assessment of surgical instrument cleanliness from sterile service departments. J Hosp Infect 2008;68:52-8. [Crossref] [PubMed]
  9. Lewis T, Griffith C, Gallo M, et al. A modified ATP benchmark for evaluating the cleaning of some hospital environmental surfaces. J Hosp Infect 2008;69:156-63. [Crossref] [PubMed]
  10. Leal Ghezzi T, Campos Corleta O. 30 Years of Robotic Surgery. World J Surg 2016;40:2550-7. [Crossref] [PubMed]
  11. Baxter RL, Baxter HC, Campbell GA, et al. Quantitative analysis of residual protein contamination on reprocessed surgical instruments. J Hosp Infect 2006;63:439-44. [Crossref] [PubMed]
  12. Alfa MJ, Fatima I, Olson N. The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels. Am J Infect Control 2013;41:249-53. [Crossref] [PubMed]
  13. Masia MD, Dettori M, Deriu GM, et al. ATP Bioluminescence for Assessing the Efficacy of the Manual Cleaning Procedure during the Reprocessing of Reusable Surgical Instruments. Healthcare (Basel) 2021;9:352. [Crossref] [PubMed]
  14. Fushimi R, Takashina M, Yoshikawa H, et al. Comparison of adenosine triphosphate, microbiological load, and residual protein as indicators for assessing the cleanliness of flexible gastrointestinal endoscopes. Am J Infect Control 2013;41:161-4. [Crossref] [PubMed]
  15. Wallace BH, Wille F, Roth K, et al. Results of performance qualification testing on clinically-used da Vinci EndoWrist instruments at hospitals in Germany. Zentralsterilisation - Central Service 2015;23:175-87.
  16. Alfa MJ, Olson N, Murray BL. Comparison of clinically relevant benchmarks and channel sampling methods used to assess manual cleaning compliance for flexible gastrointestinal endoscopes. Am J Infect Control 2014;42:e1-5. [Crossref] [PubMed]
Cite this article as: Chen A, Zou X, Tan Y, Chen Y, Ye X, Hao S. Multicenter comparative study of three “non-destructive” methods of detecting the cleanliness of the da Vinci surgical robotic instrument. Gland Surg 2021;10(12):3305-3313. doi: 10.21037/gs-21-814